Ambient radiation levels in positron emission tomography/computed tomography (PET/CT) imaging center
Creators
- 1. Universidade Federal de Minas Gerais (UFMG), Belo Horizonte, MG (Brazil)
- 2. Centro de Desenvolvimento da Tecnologia Nuclear (CDTN/CNEN-MG), Belo Horizonte, MG (Brazil)
Description
Objective: to evaluate the level of ambient radiation in a PET/CT center. Materials and methods: previously selected and calibrated TLD-100H thermoluminescent dosimeters were utilized to measure room radiation levels. During 32 days, the detectors were placed in several strategically selected points inside the PET/CT center and in adjacent buildings. After the exposure period the dosimeters were collected and processed to determine the radiation level. Results: in none of the points selected for measurements the values exceeded the radiation dose threshold for controlled area (5 mSv/ year) or free area (0.5 mSv/year) as recommended by the Brazilian regulations. Conclusion: in the present study the authors demonstrated that the whole shielding system is appropriate and, consequently, the workers are exposed to doses below the threshold established by Brazilian standards, provided the radiation protection standards are followed. (author)
Availability note (English)
Available from http://www.scielo.br/pdf/rb/v48n1/0100-3984-rb-48-01-0021.pdfAdditional details
Identifiers
Publishing Information
- Journal Title
- Radiologia Brasileira
- Journal Volume
- 48
- Journal Issue
- 1
- Journal Page Range
- p. 21-25
- ISSN
- 0100-3984
INIS
- Country of Publication
- Brazil
- Country of Input or Organization
- Brazil
- INIS RN
- 46072155
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- BUILDINGS; MEDICAL ESTABLISHMENTS; POSITRON COMPUTED TOMOGRAPHY; RADIATION DOSES; RADIATION MONITORING; RADIATION PROTECTION; STANDARDS; THERMOLUMINESCENT DOSEMETERS; THERMOLUMINESCENT DOSIMETRY
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; DIAGNOSTIC TECHNIQUES; DOSEMETERS; DOSES; DOSIMETRY; EMISSION COMPUTED TOMOGRAPHY; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; MONITORING; TOMOGRAPHY