A Study of Tissue-equivalent Compensator for 10MV X-ray and Co-60 Gamma-ray

  • 1. Dept. of Therapetic Radiology, Seoul National University, Seoul (Korea, Republic of)

Description

Authors describe some useful data when constructing tissue-equivalent compensators which would compensate tissue deficit in the treatment field of high energy electromagnetic radiation Tissue equivalent compensator is made of lucite. The ratio of compensator thickness to the thickness of tissue deficit depends on radiation energy, field size and the distance from the compensator to patient skin. When the compensator is separated from skin surface, the thickness ratio is always smaller than 1.0. This means that the larger the separation, the contribution to the total dose by means of scattered radiation from a tissue equivalent compensator is smaller. Authors propose that the thickness of lucite as tissue equivalent compensator is 0.57 times tissue deficit and the separation between compensator and skin is at least 15 m for Co-60 gamma ray and 25 cm for 10 MV X-ray.