Published July 11, 2015 | Version v1
Journal article

Development of neutron depth profiling at CMRR

Description

A neutron depth profiling (NDP) system has been developed at China Mianyang Research Reactor (CMRR) at Institute of Nuclear Physics and Chemistry (INPC), CAEP. The INPC-NDP system utilizes cold neutrons which are transported along the C1 neutron guide from the cold neutron source. It consists of a beam entrance, a target chamber, a beam stopper, and data acquisition electronics for charged particle pulse-height analysis. A 90 cm in diameter stainless steel target chamber was designed to control the positions of the sample and detector. The neutron beam intensity of 2.1×108 n cm−2 s−1 was calibrated by the Au foil activation method at the sample position. The INPC-NDP system was tested by using a Standard Reference Materials SRM-2137. The measured results agreed well with the reference values

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2015.03.058

Additional details

Identifiers

DOI
10.1016/j.nima.2015.03.058;
PII
S0168-9002(15)00383-6;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
788
Journal Page Range
p. 1-4
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Copyright
Copyright (c) 2015 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.