Thick Si(Li) coaxial detectors for registration of intermediate energy heavy ions

  • 1. St. Petersburg Nuclear Physics Institute RAS, Gatchina (Russian Federation)
  • 2. Joint Institute for Nuclear Research, Dubna (Russian Federation)

Description

Thick coaxial Si(Li) detectors are fabricated to register long-range high-ionizing reaction products with high energy resolution. It is shown that charge collection efficiency and energy resolution strongly depend on cooling temperature and magnitude of voltage applied. The test experiments have shown that coaxial detectors are able to register the intermediate energy heavy ions with 0.5-0.7% resolution. The energy losses in entrance window of the detector are <100 keV.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2008.07.148