Application of a hybrid method for efficiency calibration of a NaI detector
Creators
- 1. Nuclear Safeguards and Physical Protection Department, Egyptian Nuclear and Radiological Regulatory Authority (ENRRA), Cairo (Egypt)
- 2. Physics Department, Faculty of Science, Cairo University, Giza (Egypt)
Description
Highlights: • A novel hybrid method was extended to include other gamma–source forms. • A developed C code was used to calculate the peak efficiency numerically. • The results were validated by means of the MCNPX. • The hybrid method is a time–efficient for peak efficiency calculations. A previously proposed hybrid analytical–numerical method for efficiency calibration of a NaI detector for a point -source is extended and applied for other shapes of sources. The shapes include line, disk, and cylindrical sources of various dimensions and locations with respect to the right circular cylindrical NaI detector. The results obtained by this method have been evaluated through comparison with Monte Carlo (MC) calculations. A relative maximum difference of 3.5% is found between the two methods.
Availability note (English)
Available from http://dx.doi.org/10.1016/j.apradiso.2021.109632Additional details
Identifiers
- DOI
- 10.1016/j.apradiso.2021.109632;
- PII
- S0969804321000452;
Publishing Information
- Journal Title
- Applied Radiation and Isotopes
- Journal Volume
- 171
- Journal Page Range
- vp.
- ISSN
- 0969-8043
- CODEN
- ARISEF
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 54076267
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S97: MATHEMATICAL METHODS AND COMPUTING;
- Descriptors DEI
- C CODES; CALIBRATION; CYLINDRICAL CONFIGURATION; MONTE CARLO METHOD; POINT SOURCES
- Descriptors DEC
- CALCULATION METHODS; COMPUTER CODES; CONFIGURATION; RADIATION SOURCES
Optional Information
- Copyright
- Copyright (c) 2021 Elsevier Ltd. All rights reserved.