Published August 1992 | Version v1
Journal article

Deuterated liquid scintillator (NE230) as a fast neutron detector for cold-fusion and other research

  • 1. Michigan Univ., Ann Arbor, MI (United States). Dept. of Physics

Description

Unlike conventional liquid scintillators (e.g. NE213) deuterated scintillators (NE230) can provide, without time-of-flight, usable neutron energy spectra for neutrons En ≥ 1 MeV. In this paper, the authors assemble and calibrate several of these detectors and have used them in experiments searching for d + d fusion-neutron emission (En doteq 2.5 MeV) in Pd + D2O electrolysis and Ti + D2 adsorption at LN temperatures. The detectors yield direct fusion-neutron spectra and set limits in the Pd + D and Ti + D experiments of < 7 x 10-24 fusion n/s/dd pair and < 3 x 10-24 fusion n/s/dd pair respectively. In addition, these detectors have been used for in-beam accelerator experiments requiring coincidence between fast neutrons and scattered ions

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
39
Journal Issue
4
Journal Page Range
p. 532-535.
ISSN
0018-9499
CODEN
IETNAE

Conference

Title
1991 Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
Dates
2-9 Nov 1991.
Place
Santa Fe, NM (United States).

Optional Information

Secondary number(s)
CONF-911106--.