Published October 1992 | Version v1
Journal article

Gamma ray spectrometer readout with linear CCD sensor

  • 1. Lawrence Berkeley Lab., CA (United States)
  • 2. Los Alamos National Lab., NM (United States)

Description

In this paper, a high resolution linear TV camera is described for readout of phosphors illuminated by gated microchannel plate image intensifiers. The 2592-element array clocking scheme allows for simultaneous readout of even and odd pixels combined into a single photoelement. Improved signal noise ratio results, exceeding 1000:1 for a single video line readout. The camera can be reset at any time and unwanted photocharge dumped. Integration time is adjustable to match phosphor decay time and clock rates exceeding 20 MHz allow for a fast readout. Measurements including resolution, responsitivity, and dynamic range of the camera are presented

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
39
Journal Issue
5
Journal Page Range
p. 1336-1340.
ISSN
0018-9499
CODEN
IETNAE

Conference

Title
1991 Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
Dates
2-9 Nov 1991.
Place
Santa Fe, NM (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
24047982
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
CHARGE-COUPLED DEVICES; GAMMA SPECTROMETERS; IMAGE INTENSIFIERS; IMAGE PROCESSING; PERFORMANCE; PHOSPHORS; TELEVISION CAMERAS
Descriptors DEC
CAMERAS; MEASURING INSTRUMENTS; SEMICONDUCTOR DEVICES; SPECTROMETERS

Optional Information

Secondary number(s)
CONF-911106--.