Published October 1992
| Version v1
Journal article
Gamma ray spectrometer readout with linear CCD sensor
- 1. Lawrence Berkeley Lab., CA (United States)
- 2. Los Alamos National Lab., NM (United States)
Description
In this paper, a high resolution linear TV camera is described for readout of phosphors illuminated by gated microchannel plate image intensifiers. The 2592-element array clocking scheme allows for simultaneous readout of even and odd pixels combined into a single photoelement. Improved signal noise ratio results, exceeding 1000:1 for a single video line readout. The camera can be reset at any time and unwanted photocharge dumped. Integration time is adjustable to match phosphor decay time and clock rates exceeding 20 MHz allow for a fast readout. Measurements including resolution, responsitivity, and dynamic range of the camera are presented
Additional details
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 39
- Journal Issue
- 5
- Journal Page Range
- p. 1336-1340.
- ISSN
- 0018-9499
- CODEN
- IETNAE
Conference
- Title
- 1991 Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
- Dates
- 2-9 Nov 1991.
- Place
- Santa Fe, NM (United States).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 24047982
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CHARGE-COUPLED DEVICES; GAMMA SPECTROMETERS; IMAGE INTENSIFIERS; IMAGE PROCESSING; PERFORMANCE; PHOSPHORS; TELEVISION CAMERAS
- Descriptors DEC
- CAMERAS; MEASURING INSTRUMENTS; SEMICONDUCTOR DEVICES; SPECTROMETERS
Optional Information
- Secondary number(s)
- CONF-911106--.