Published May 1983
| Version v1
Journal article
Cantilever and capacitor technique for measuring dilatation
Creators
- 1. Argonne National Laboratory, Solid State Science Division, Argonne, Illinois 60439
Description
The relationship of EerNisse's technique for measuring small dilatations caused by irradiation with short-range particles, which utilizes a metallized thin plate mounted as a cantilever below whose free end an electrode is placed (forming a capacitor), to a photoelastic technique and to an interferometric technique are derived. The effects of stray capacitance, the fringing field of the capacitor, the clamping stress on the cantilever plate, the electrical resistance of the metallic coating, the charging of the tank circuit of which the capacitor is an element, the flange bolting stress, and the beam heating are assessed, and examples of the manner in which they contaminate the data are given
Additional details
Publishing Information
- Journal Title
- Rev. Sci. Instrum.
- Journal Volume
- 54
- Journal Issue
- 5
- Series
- Rev. Sci. Instrum.
- Journal Page Range
- 544-551
- ISSN
- 0034-6748
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 14784130
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CAPACITORS; EXPANSION; IRRADIATION; MATHEMATICAL MODELS; MEASURING INSTRUMENTS; PHOTOELASTICITY; PLATES; RANGE; SPECIFICATIONS; THICKNESS
- Descriptors DEC
- DIMENSIONS; ELASTICITY; ELECTRICAL EQUIPMENT; ENERGY STORAGE SYSTEMS; EQUIPMENT; MECHANICAL PROPERTIES