Estimated fetal radiation dose from multislice CT studies
- 1. Mayo Clinic and Foundation, Rochester, MN (USA)
Description
It is often necessary to estimate the dose of radiation to a fetus from a series of CT scans. To assist in making this process easier and more accurate, we measured fetal doses in an adult anthropomorphic phantom for four CT scanners: Picker 1200, Siemens DRH, GE 9800, and GE 8800. Measurements were made at four kilovoltages (100, 120, 130, and 140 kVp), for 2-, 5-, 8-, and 10-mm thicknesses, for two scanning angles (360 degrees and 398 degrees) and for two patient orientations (prone and supine). The fetal-dose estimates are based on the CT dose index measured by using a pencil ionization chamber at the center position of a 16-cm-diameter cylindrical phantom. Comparison with data from other investigators shows reasonable agreement when phantom differences, X-ray tube output, and equipment calibration are considered
Additional details
Publishing Information
- Journal Title
- AJR, American Journal of Roentgenology
- Journal Volume
- 154
- Journal Issue
- 1
- Series
- AJR, Am. J. Roentgenol.
- Journal Page Range
- 185-190
- ISSN
- 0361-803X
- CODEN
- AAJRD
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 21043199
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; DIAGNOSIS; FETUSES; IONIZATION CHAMBERS; MAN; PHANTOMS; PREGNANCY; RADIATION DOSES; X RADIATION
- Descriptors DEC
- ANIMALS; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MAMMALS; MEASURING INSTRUMENTS; MOCKUP; PRIMATES; RADIATION DETECTORS; RADIATIONS; STRUCTURAL MODELS; TOMOGRAPHY; VERTEBRATES