Published January 1990 | Version v1
Journal article

Estimated fetal radiation dose from multislice CT studies

  • 1. Mayo Clinic and Foundation, Rochester, MN (USA)

Description

It is often necessary to estimate the dose of radiation to a fetus from a series of CT scans. To assist in making this process easier and more accurate, we measured fetal doses in an adult anthropomorphic phantom for four CT scanners: Picker 1200, Siemens DRH, GE 9800, and GE 8800. Measurements were made at four kilovoltages (100, 120, 130, and 140 kVp), for 2-, 5-, 8-, and 10-mm thicknesses, for two scanning angles (360 degrees and 398 degrees) and for two patient orientations (prone and supine). The fetal-dose estimates are based on the CT dose index measured by using a pencil ionization chamber at the center position of a 16-cm-diameter cylindrical phantom. Comparison with data from other investigators shows reasonable agreement when phantom differences, X-ray tube output, and equipment calibration are considered

Additional details

Publishing Information

Journal Title
AJR, American Journal of Roentgenology
Journal Volume
154
Journal Issue
1
Series
AJR, Am. J. Roentgenol.
Journal Page Range
185-190
ISSN
0361-803X
CODEN
AAJRD