Published 2014 | Version v1
Journal article

Studying spatial resolution of CZT detectors using sub-pixel positioning for SPECT

  • 1. CEA Grenoble, DRT-LETI, F-38054 Grenoble 9, (France)

Description

CZT detectors are the basic building block of a variety of new SPECT systems. Their modularity allows adapting system architecture to specific applications such as cardiac, breast, brain or small animal imaging. In semiconductors, a high number of electron-hole pairs is produced by a single interaction. This direct conversion process allows better energy and spatial resolutions than usual scintillation detectors based on NaI(Tl). However, it remains often unclear if SPECT imaging can really benefit of that performance gain. We investigate the system performance of a detection module, which is based on 5 mm thick CZT with a segmented anode having a 2.5 mm pitch by simulation and experimentation. This pitch allows an easy assembly of the crystal on the readout board and limits the space occupied by electronics without significantly degrading energy and spatial resolution. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1109/TNS.2014.2356293

Additional details

Identifiers

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
61
Journal Issue
no.5
Journal Page Range
p. 2559-2566
ISSN
0018-9499

Optional Information

Notes
26 refs.