Studying spatial resolution of CZT detectors using sub-pixel positioning for SPECT
- 1. CEA Grenoble, DRT-LETI, F-38054 Grenoble 9, (France)
Description
CZT detectors are the basic building block of a variety of new SPECT systems. Their modularity allows adapting system architecture to specific applications such as cardiac, breast, brain or small animal imaging. In semiconductors, a high number of electron-hole pairs is produced by a single interaction. This direct conversion process allows better energy and spatial resolutions than usual scintillation detectors based on NaI(Tl). However, it remains often unclear if SPECT imaging can really benefit of that performance gain. We investigate the system performance of a detection module, which is based on 5 mm thick CZT with a segmented anode having a 2.5 mm pitch by simulation and experimentation. This pitch allows an easy assembly of the crystal on the readout board and limits the space occupied by electronics without significantly degrading energy and spatial resolution. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1109/TNS.2014.2356293Additional details
Identifiers
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 61
- Journal Issue
- no.5
- Journal Page Range
- p. 2559-2566
- ISSN
- 0018-9499
INIS
- Country of Publication
- United States
- Country of Input or Organization
- France
- INIS RN
- 48091873
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANODES; GAMMA RADIATION; READOUT SYSTEMS; SEMICONDUCTOR DETECTORS; SIMULATION; SPATIAL RESOLUTION
- Descriptors DEC
- ELECTRODES; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATION DETECTORS; RADIATIONS; RESOLUTION
Optional Information
- Notes
- 26 refs.