Published December 11, 1997 | Version v1
Journal article

A time-of-flight spectrometer with pulse-shape discrimination for the measurement of double-differential charged-particle emission cross sections

  • 1. Osaka Univ. (Japan). Dept. of Nuclear Engineering
  • 2. Free Electron Laser Lab., Osaka (Japan)
  • 3. Sumitomo Heavy Industry, Saijo, Ehime (Japan)
  • 4. Mitsubishi Electric Corp., Amagasaki, Hyogo (Japan). Itami Works
  • 5. Toshiba Co., Tokyo (Japan)

Description

Based on the two-dimensional analysis of energy and time-of-flight spectra, combined with the pulse-shape discrimination technique using a CsI(Tl) scintillation detector, a charged-particle spectrometer was developed to measure double-differential charged-particle emission cross sections with 14 MeV incident neutrons. Using this spectrometer, proton emission cross sections for Fe and α-particle emission cross sections for Be were measured to confirm the basic performance of the spectrometer. The measured data were compared with other experimental results, theoretical calculations and evaluations. Also, a model calculation based on kinematics for α-particle emission from Be was presented. (orig.)

Additional details

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
401
Journal Issue
1
Journal Page Range
p. 93-103.
ISSN
0168-9002
CODEN
NIMAER

Optional Information

Notes
13 refs.