Published June 1, 1984
| Version v1
Journal article
Improved technique for recovery and measurement of polonium-210 from environmental materials
Creators
- 1. Melbourne Univ., Parkville (Australia). School of Chemistry
Description
Polonium-210 obtained in acidic solution by digestion of sediment or tissue samples is efficiently plated onto one side of a silver disc for measurement of α-activity. Optimum conditions for solutions obtained from environmental samples are pH 1.5-2 in the presence of citrate and ascorbic acid, with the solution at 800C and the silver disc rotated at 100-150 radian s-1. Observed recoveries using a polonium-208 tracer are 97.5 +- 2.5% for sediments and 99.3 +- 1.3% for tissues. (Auth.)
Additional details
Publishing Information
- Journal Title
- Anal. Chim. Acta
- Journal Volume
- 160
- Series
- Anal. Chim. Acta.
- Journal Page Range
- 69-77
- ISSN
- 0003-2670
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 15070441
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- ALPHA DETECTION; CONTAMINATION; DEPOSITION; ENVIRONMENT; LI-DRIFTED SI DETECTORS; PH VALUE; POLONIUM 210; SEDIMENTS; TISSUES
- Descriptors DEC
- ALPHA DECAY RADIOISOTOPES; BODY; CHARGED PARTICLE DETECTION; DAYS LIVING RADIOISOTOPES; DETECTION; EVEN-EVEN NUCLEI; HEAVY NUCLEI; ISOTOPES; LI-DRIFTED DETECTORS; MEASURING INSTRUMENTS; NUCLEI; POLONIUM ISOTOPES; RADIATION DETECTION; RADIATION DETECTORS; RADIOISOTOPES; SEMICONDUCTOR DETECTORS; SI SEMICONDUCTOR DETECTORS