Published June 2001 | Version v1
Journal article

Inter-comparison of geometrical track parameters and depth dependent track etch rates measured for Li-7 ions in two types of CR-39

Description

The depth dependent track etch rates of two types of CR-39, TASTRAK and BARYOTRAK, for Li-7 ions with incident energies of 4.82, 6.75 and 10.77 MeV have been evaluated from track length measurements. The pit lengths versus the etching time obtained by three different laboratories have the same trend, but did not agree completely. The differences can be attributed to an error margin of less than one decimal point in the etching temperature. Significant deviations were also present for the assessed track etch rates. Most of the deviation was attributed to the derivation step of the growth curves. A sufficient amount of data and reasonably smooth growth curves were required near the Bragg peak in order to make a precise estimation

Additional details

Identifiers

PII
S1350448701001172;

Publishing Information

Journal Title
Radiation Measurements
Journal Volume
34
Journal Issue
1-6
Journal Page Range
p. 37-43
ISSN
1350-4487
CODEN
RMEAEP

Optional Information

Copyright
Copyright (c) 2001 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.