Can a commercial gel dosimetry system be used to verify stereotactic spinal radiotherapy treatment dose distributions?
Creators
- 1. School of Chemistry, Physics, and Mechanical Engineering, Queensland University of Technology, Brisbane (Australia)
- 2. Royal Brisbane and Women's Hospital, Brisbane (Australia)
- 3. Genesis Cancer Care Queensland, Brisbane (Australia)
- 4. School of Science, RMIT University, Melbourne (Australia)
Description
This study investigated the use of the TruView xylenol-orange-based gel and VISTA optical CT scanner (both by Modus Medical Inc, London, Canada), for use in verifying the accuracy of planned dose distributions for hypo-fractionated (stereotactic) vertebral treatments. Gel measurements were carried out using three stereotactic vertebral treatments and compared with planned doses calculated using the Eclipse treatment planning system (Varian Medical Systems, Palo Alto, USA) as well as with film measurements made using Gafchromic EBT3 film (Ashland Inc, Covington, USA), to investigate the accuracy of the gel system. The gel was calibrated with reference to a moderate-dose gradient region in one of the gel samples. Generally, the gel measurements were able to approximate the close agreement between the doses calculated by the treatment planning system and the doses measured using film (which agreed with each other within 2%), despite lower resolution and bit depth. Poorer agreement was observed when the dose delivered to the gel exceeded the range of doses delivered in the calibration region. This commercial gel dosimetry system may be used to verify hypo-fractionated treatments of vertebral targets, although separate gel calibration measurements are recommended. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/1742-6596/847/1/012071Additional details
Identifiers
Publishing Information
- Journal Title
- Journal of Physics. Conference Series (Online)
- Journal Volume
- 847
- Journal Issue
- 1
- Journal Page Range
- [5 p.]
- ISSN
- 1742-6596
Conference
- Title
- 9. international conference on 3D radiation dosimetry
- Dates
- 7-10 Nov 2016
- Place
- Galveston, TX (United States)
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 49012412
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; CALIBRATION; COMPARATIVE EVALUATIONS; COMPUTERIZED TOMOGRAPHY; DOSIMETRY; FILMS; PLANNING; POLYMER GEL DOSEMETERS; POLYMER GEL DOSIMETRY; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; RADIOTHERAPY; XYLENOL ORANGE; XYLENOLS
- Descriptors DEC
- AROMATICS; CHEMICAL DOSEMETERS; DIAGNOSTIC TECHNIQUES; DOSEMETERS; DOSES; DOSIMETRY; DYES; EVALUATION; HYDROCARBONS; HYDROXY COMPOUNDS; INDICATORS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; ORGANIC COMPOUNDS; PHENOLS; RADIOLOGY; THERAPY; TOMOGRAPHY