Published March 2005 | Version v1
Report Open

Calculations of an HPGe detector peak efficiency curve up to 11 MeV with EGS4 and GEANT4

  • 1. Nagoya Univ., Graduate School of Engineering, Nagoya, Aichi (Japan)
  • 2. Nagoya Univ., Radioisotope Research Center, Nagoya, Aichi (Japan)
  • 3. Kyoto Univ., Research Reactor Institute, Kumatori, Osaka (Japan)

Description

We determined an efficiency curve of an HPGe detector in the energy range of 0.3 - 11 MeV within 0.5% accuracy with the measured data of the 14N(n,γ)15N reaction and the aid of the Monte Carlo cords EGS4 and GEANT4. (author)

Files

36113850.pdf

Files (303.6 kB)

Name Size Download all
md5:7dc69782b269aa506263c59d8da78780
303.6 kB Preview Download
Part of:
Proceedings of the 2004 symposium on nuclear data

Additional details

Publishing Information

Imprint Title
Proceedings of the 2004 symposium on nuclear data
Imprint Pagination
268 p.
Journal Page Range
p. 216-221
Report number
JAERI-Conf--2005-003

Conference

Title
2004 symposium on nuclear data
Dates
11-12 Nov 2004
Place
Tokai, Ibaraki (Japan)

Optional Information

Notes
4 refs., 9 figs.; This record replaces 36085539
Secondary number(s)
INDC(JPN)--195/U