The structural and bio-corrosion barrier performance of Mg-substituted fluorapatite coating on 316L stainless steel human body implant

  • 1. Biomaterials Group, Department of Metallurgy and Materials Engineering, Iran University of Science and Technology, Tehran, 16844 (Iran, Islamic Republic of)
  • 2. Dental Materials Research Center, Isfahan University of Medical Sciences, Isfahan (Iran, Islamic Republic of)
  • 3. Biomaterials Research Group, Department of Materials Engineering, Isfahan University of Technology, Isfahan, 8415683111 (Iran, Islamic Republic of)

Description

In this study, Mg-substituted fluorapatite coatings were deposited on medical grade AISI 316L stainless steel via sol–gel dip coating method. Phase composition, crystallite size and degree of crystallinity of the obtained coatings were evaluated by X-ray diffraction (XRD) analysis. Fourier transform infrared (FTIR) spectroscopy was also used to evaluate functional groups of the obtained coatings. The surface morphology and cross-section of the final coatings were studied using scanning electron microscopy (SEM) and energy dispersive X-ray (EDX) spectroscopy was used to determine elemental chemical composition of the obtained coatings. In order to determine and compare the corrosion behavior of uncoated and Mg-substituted fluorapatite coated 316L stainless steel, electrochemical potentiodynamic polarization tests were performed in physiological solutions at 37 ± 1 °C. Moreover, the released metallic ions from uncoated and coated substrates were measured by inductively coupled plasma-optical emission spectrometry (ICP-OES) within 2 months of immersing in Ringer's solution at 36.5 ± 1 °C as an indication of biocompatibility. The results showed that fluoride and magnesium were successfully incorporated into apatite lattice structure and the prepared coatings were nanostructured with crystallinity of about 70%. Obtained coatings were totally crack-free and uniform and led to decrease in corrosion current densities of 316L stainless steel in physiological solutions. In addition, coated sample released much less ions such as Fe, Cr and Ni in physiological media. Therefore, it was concluded that Mg-substituted fluorapatite coatings could improve the corrosion resistance and biocompatibility of 316L stainless steel human body implants.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apsusc.2013.10.029

Additional details

Identifiers

DOI
10.1016/j.apsusc.2013.10.029;
PII
S0169-4332(13)01876-X;

Publishing Information

Journal Title
Applied Surface Science
Journal Volume
288
Journal Page Range
p. 331-340