Spatially resolved X-ray fluorescence analysis with a pyroelectric X-ray emitter
- 1. Kyoto Univ., Dept. of Materials Science and Engineering, Kyoto (Japan)
Description
A portable instrument for two-dimensional X-ray fluorescence imaging was assembled with an X-ray source using a pyroelectric crystal, which was driven by a 9-V dry battery, a Si-PIN detector, a slit, and pulse motors. Line scanning for a mug and a knife-edge-scan of an iron sheet were carried out using this spectrometer. The sensitivity of the spectrometer was sufficient for elemental analysis of a mug using a 1 mm2 slit, and several elements, such as Co, Ni, Zn, Pb and Zr, were detected. The estimated spatial resolution using a 0.8-mm pinhole was 3.5 mm. (author)
Additional details
Publishing Information
- Journal Title
- Analytical Sciences
- Journal Volume
- 24
- Journal Issue
- 7
- Journal Page Range
- p. 847-850
- ISSN
- 0910-6340
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 39108971
- Subject category
- S37: INORGANIC, ORGANIC, PHYSICAL AND ANALYTICAL CHEMISTRY;
- Descriptors DEI
- BEAM OPTICS; IMAGE PROCESSING; PYROELECTRIC EFFECT; SENSITIVITY; SPATIAL RESOLUTION; X-RAY DETECTION; X-RAY FLUORESCENCE ANALYSIS; X-RAY SOURCES; X-RAY SPECTRA; X-RAY SPECTROMETERS
- Descriptors DEC
- CHEMICAL ANALYSIS; DETECTION; MEASURING INSTRUMENTS; NONDESTRUCTIVE ANALYSIS; PROCESSING; RADIATION DETECTION; RADIATION SOURCES; RESOLUTION; SPECTRA; SPECTROMETERS; X-RAY EMISSION ANALYSIS
Optional Information
- Notes
- 21 refs., 6 figs.