Published July 1980
| Version v1
Journal article
Optimization of the trochoidal electron monochromator
Creators
- 1. Chemistry Department, University of Maryland, College Park, Maryland 20742
Description
The trochoidal electron monochromator employs an axial magnetic field and a transverse electric field. Many aspects of its performance can be described by electron trajectory equations which include the voltage drop across the entrance aperture. A beam shear is predicted, which is a function of a parameter tantheta. Optimum performance is achieved by matching the exit aperture to the sheared beam. tantheta establishes a relationship between the field magnitudes and the dimensions of the monochromator. Estimates of appropriate values for these parameters can be made, with the final optimum values being determined by a combination of practical and empirical considerations. Finally, the use of the instrument as an energy analyzer is described
Additional details
Publishing Information
- Journal Title
- Rev. Sci. Instrum.
- Journal Volume
- 51
- Journal Issue
- 7
- Series
- Rev. Sci. Instrum.
- Journal Page Range
- 944-950
- ISSN
- 0034-6748
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 11559962
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- APERTURES; ELECTRIC FIELDS; ELECTRIC POTENTIAL; ELECTRON SPECTROMETERS; EQUATIONS; MAGNETIC FIELDS; MONOCHROMATORS; OPTIMIZATION; TRAJECTORIES
- Descriptors DEC
- MEASURING INSTRUMENTS; OPENINGS; SPECTROMETERS