Published February 1983
| Version v1
Journal article
Artifacts in iodine-131 renal images due to coincidence summing of technetium-99m photons
- 1. Department of Radiology, Duke University Medical Center, Durham, NC
Description
Technetium-99m gamma camera images of kidneys produced with an iodine-131 energy window showed ghosts over and between the kidneys and between the kidneys and the bladder. The cause was coincidence summing at 281 keV due to a high /sup 99m/Tc counting rate. In older model cameras careful peaking with a proper energy window reduces this effect; more recently developed cameras with pileup rejection circuits do not have this problem
Additional details
Publishing Information
- Journal Title
- Radiology
- Journal Volume
- 146
- Journal Issue
- 2
- Series
- Radiology.
- Journal Page Range
- 505-507
- ISSN
- 0033-8419
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 14779541
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BLADDER; EFFICIENCY; ERRORS; GAMMA CAMERAS; IMAGE PROCESSING; IMAGES; IODINE 131; ISOMERIC NUCLEI; KIDNEYS; PHOTONS; TECHNETIUM 99; USA
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; BODY; CAMERAS; DAYS LIVING RADIOISOTOPES; DEVELOPED COUNTRIES; ELEMENTARY PARTICLES; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; IODINE ISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MASSLESS PARTICLES; NORTH AMERICA; NUCLEI; ODD-EVEN NUCLEI; ORGANS; RADIOISOTOPES; TECHNETIUM ISOTOPES; URINARY TRACT; YEARS LIVING RADIOISOTOPES