Published July 1, 1984 | Version v1
Journal article

Number of photoelectrons from a photomultiplier cathode coupled with a NaI(Tl) scintillator

  • 1. National Lab. for High Energy Physics, Oho, Ibaraki (Japan)
  • 2. Saitama Coll. of Health, Urawa (Japan)

Description

Absolute numbers of photoelectrons from several PMT photocathodes coupled with NaI(Tl) scintillators were measured for the several energies of gamma-rays. Typically, 6.85 x 103 photoelectrons were obtained with a resolution of 13% for 662 keV gamma-rays of 137Cs. (orig.)

Additional details

Publishing Information

Journal Title
Nucl. Instrum. Methods Phys. Res., Sect. A
Journal Volume
224
Journal Issue
1/2
Series
CODEN: NIMRD.;Nucl. Instrum. Methods Phys. Res., Sect. A.
Journal Page Range
331-334
ISSN
0167-5087