Published 2007 | Version v1
Journal article

Model for the response of radiographic films and implications for quality assurance

  • 1. Klinik und Poliklinik fuer Strahlentherapie und Radioonkologie, Onkologisches Zentrum, Universitaetsklinikum Hamburg-Eppendorf (Germany)

Description

The present paper describes the development of a model linking the optical density of a radiographic film to the applied dose. This model takes into account the kind of bremsstrahlung used in radiotherapy, and uses fewer parameters compared with the models developed so far. The measurements showed an influence of the field size on the model parameters. Monte Carlo simulations showed that different field sizes and depths in solid water can change the spectral distributions. The spectrum hardens with decreasing field size and increasing depth in phantom material. This has an influence on the model parameters. Nevertheless, the model can still be used to describe the relation between optical density and applied dose for films that are irradiated with fields of different size. A dose-rate dependence was not observed. The size and form of the fields used in calibration and application can therefore cause a systematic error in verification with film. (orig.)

Additional details

Additional titles

Original title (German)
Modell fuer das Ansprechen radiographischer Filme und Folgerungen fuer die Qualitaetssicherung

Publishing Information

Journal Title
Zeitschrift fuer Medizinische Physik
Journal Volume
17
Journal Issue
3
Journal Page Range
p. 197-204
ISSN
0939-3889

INIS

Country of Publication
Germany
Country of Input or Organization
Germany
INIS RN
39005725
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; FILMS; IMAGE PROCESSING; IMAGES; QUALITY ASSURANCE
Descriptors DEC
DIAGNOSTIC TECHNIQUES; MEDICINE; NUCLEAR MEDICINE; PROCESSING; RADIOLOGY