Published 1975 | Version v1
Report Open

Optimization attempts for time resolution in fast-slow coincidence systems

Description

A systematic investigation to obtain good time resolutions with proper shape of the resolution curve in fast-slow coincidence systems is described. Measuring the output time-walk as a function of the input signal amplitude for fast timing discriminators, it was possible to choose an input signal amplitude range resulting in small time uncertainty and thus to achieve a good time resolution almost independent of the dynamic range. With an additional fast gating line a more gaussian shape of the resolution curve with less pronounced tails on both sides could be obtained

Availability note (English)

MF available from INIS under the Report Number.

Abstract (French)

Une analyse systematique des courbes de resolution prompte obtenues avec un systeme de coincidence lent-rapide a ete effectuee. Des fluctuations en temps importantes de l'impulsion de sortie des discriminateurs rapides suivant l'amplitude de l'impulsion d'entree ont ete observees. L'ajustage de cette amplitude et l'adjonction d'un circuit de coincidence rapide autorisant la prise en compte de l'impulsion de depart du convertisseur temps-amplitude ont permis d'ameliorer la largeur et la symetrie de la courbe de resolution prompte

Files

7246236.pdf

Files (274.5 kB)

Name Size Download all
md5:fb4545830395cb297013a7c5dc767fb2
274.5 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
6 p.
Report number
CRN-CNPA--75-25

INIS

Country of Publication
France
Country of Input or Organization
France
INIS RN
7246236
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
COINCIDENCE CIRCUITS; OPTIMIZATION; TIME RESOLUTION
Descriptors DEC
ELECTRONIC CIRCUITS; RESOLUTION; TIMING PROPERTIES