Performance evaluation of the whole-body PET scanner ECAT EXACT HR+
- 1. German Cancer Research Center, Heidelberg (Germany)
Description
The performance parameters of the whole-body PET scanner ECAT EXACT HR+ were determined following the standard proposed by the International Electrotechnical Commission (IEC). The tests were expanded by some measurements concerning the accuracy of the correction algorithms and the geometric fidelity of the reconstructed images. The scanner consists of 32 rings, each with 576 BGO detectors (4.05 x 4.39 x 30 mm3) covering an axial field-of-view of 15.5 cm and a patient port of 56.2 cm. The transaxial resolution in the 2D (3D) mode is 4.5 (4.3) mm at the center. It increases to 8.9 (8.3) mm radially and to 5.8 (5.2) mm tangentially at a radial distance of r = 20 cm. The average axial resolution varies between 4.9 (4.1) mm FWHM at the center and 8.8 (8.1) mm at r = 20 cm. The system sensitivity for true events is 5.85 (26.4) cps/Bq/ml (measured with a 20 cm cylinder phantom). The 50% dead-time losses where reached for a true event count rate of 286 (500) kcps at an activity concentration of 74 (25) kBq/ml. The system scatter fraction is 0.24 (0.35). The correction algorithms work reliable, except for the 3D attenuation correction. The ECAT EXACT HR+ has a good and nearly isotropic spatial resolution. Due to the small detector elements, however, it has a low slice sensitivity which is a limiting factor for image quality
Additional details
Publishing Information
- Publisher
- IEEE Service Center.
- Imprint Place
- Piscataway, NJ (United States)
- Imprint Title
- 1996 IEEE nuclear science symposium - conference record. Volumes 1, 2 and 3
- Imprint Pagination
- 2138 p.
- Journal Page Range
- p. 1270-1274.
Conference
- Title
- Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
- Dates
- 2-9 Nov 1996.
- Place
- Anaheim, CA (United States).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 28068746
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ALGORITHMS; DATA ACQUISITION SYSTEMS; IMAGE PROCESSING; PERFORMANCE TESTING; POSITRON COMPUTED TOMOGRAPHY; SPATIAL RESOLUTION
- Descriptors DEC
- COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; RESOLUTION; TESTING; TOMOGRAPHY
Optional Information
- Secondary number(s)
- CONF-961123--.