Performance evaluation of the whole-body PET scanner ECAT EXACT HR+

  • 1. German Cancer Research Center, Heidelberg (Germany)

Description

The performance parameters of the whole-body PET scanner ECAT EXACT HR+ were determined following the standard proposed by the International Electrotechnical Commission (IEC). The tests were expanded by some measurements concerning the accuracy of the correction algorithms and the geometric fidelity of the reconstructed images. The scanner consists of 32 rings, each with 576 BGO detectors (4.05 x 4.39 x 30 mm3) covering an axial field-of-view of 15.5 cm and a patient port of 56.2 cm. The transaxial resolution in the 2D (3D) mode is 4.5 (4.3) mm at the center. It increases to 8.9 (8.3) mm radially and to 5.8 (5.2) mm tangentially at a radial distance of r = 20 cm. The average axial resolution varies between 4.9 (4.1) mm FWHM at the center and 8.8 (8.1) mm at r = 20 cm. The system sensitivity for true events is 5.85 (26.4) cps/Bq/ml (measured with a 20 cm cylinder phantom). The 50% dead-time losses where reached for a true event count rate of 286 (500) kcps at an activity concentration of 74 (25) kBq/ml. The system scatter fraction is 0.24 (0.35). The correction algorithms work reliable, except for the 3D attenuation correction. The ECAT EXACT HR+ has a good and nearly isotropic spatial resolution. Due to the small detector elements, however, it has a low slice sensitivity which is a limiting factor for image quality

Additional details

Publishing Information

Publisher
IEEE Service Center.
Imprint Place
Piscataway, NJ (United States)
Imprint Title
1996 IEEE nuclear science symposium - conference record. Volumes 1, 2 and 3
Imprint Pagination
2138 p.
Journal Page Range
p. 1270-1274.

Conference

Title
Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
Dates
2-9 Nov 1996.
Place
Anaheim, CA (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
28068746
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
ALGORITHMS; DATA ACQUISITION SYSTEMS; IMAGE PROCESSING; PERFORMANCE TESTING; POSITRON COMPUTED TOMOGRAPHY; SPATIAL RESOLUTION
Descriptors DEC
COMPUTERIZED TOMOGRAPHY; EMISSION COMPUTED TOMOGRAPHY; RESOLUTION; TESTING; TOMOGRAPHY

Optional Information

Secondary number(s)
CONF-961123--.