Published December 2019 | Version v1
Journal article

Measurements of temperature and heat of phase transformation of pure silicon by using differential scanning calorimetry

  • 1. Foundry Research Institute (Poland)

Description

Differential scanning calorimetry (DSC) technique has been applied for the experimental determination of temperature and heat of phase transition of pure silicon (7 N) during heating and cooling cycles at the rate of 10 K min−1. The measurements were carried out in the temperature range of 25–1450 °C in a flow gas atmosphere (Ar, 99.9992%) using three types of crucibles made of alumina, h-BN and alumina covered with h-BN coating. The following characteristics were estimated from DSC curves: melting point of silicon—1414 °C, the heat of fusion—1826 J g−1 and the heat of solidification—1654 J g−1. It was found that the silicon evaporation phenomenon accompanying the tests had no effect on the measurements of temperature during solid-to-liquid and liquid-to-solid transformations and on the measurement of the latent heat of fusion. The effect of crucible type on the DSC measurements is discussed.

Additional details

Identifiers

Publishing Information

Journal Title
Journal of Thermal Analysis and Calorimetry
Journal Volume
138
Journal Issue
6
Journal Page Range
p. 4215-4221
ISSN
1388-6150

Optional Information

Copyright
Copyright (c) 2019 The Author(s)