Characteristics study on the APD-based front-end readout for CsI(Tl) detector
- 1. Nuclear Technology Key Laboratory of Earth Science, Chengdu University of Technology, Chengdu (China)
Description
Abstract Background: The avalanche photodiode (APD) has the advantages of small size, high detection efficiency, internal gain and insensitivity to magnet fields. But the internal gain and the signal-noise ratio (SNR) of the output signal of pulse are affected by bias voltage and temperature. Purpose: This study aims to test the effect of bias voltage and temperature on the APD-based front-end readout of scintillation detector. Methods: The avalanche photodiode is employed as the photodetector for CsI(Tl) scintillation, and equiped a charge sensitive preamplifier of low noise to form a probe. The energy resolution of scintillation detector with this probe was tested on a 137Cs radiation source under different bias voltage and temperature. Results: The experimental results show that the energy resolution of the detection system is affected by both the bias voltage and the temperature. When the bias voltage at both ends of the APD is 370 V, the energy resolution is 4.98% for the gamma ray with the energy of 662 keV at the room temperature. Conclusion: The energy resolution will increase with the decrease of temperature within the range of -20℃ to 40℃. (authors)
Additional details
Identifiers
Publishing Information
- Journal Title
- Nuclear Techniques
- Journal Volume
- 39
- Journal Issue
- 10
- Journal Page Range
- [5 p.]
- ISSN
- 0253-3219
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 51087802
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CESIUM 137; EFFICIENCY; ELECTRIC POTENTIAL; ENERGY RESOLUTION; GAMMA RADIATION; KEV RANGE 100-1000; NOISE; PHOTODETECTORS; PHOTODIODES; PREAMPLIFIERS; PULSES; RADIATION SOURCES; READOUT SYSTEMS; SCINTILLATION COUNTERS; SCINTILLATIONS; SIGNALS; SIGNAL-TO-NOISE RATIO; TEMPERATURE RANGE 0273-0400 K
- Descriptors DEC
- AMPLIFIERS; BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CESIUM ISOTOPES; DIMENSIONLESS NUMBERS; ELECTROMAGNETIC RADIATION; ELECTRONIC EQUIPMENT; ENERGY RANGE; EQUIPMENT; INTERMEDIATE MASS NUCLEI; IONIZING RADIATIONS; ISOTOPES; KEV RANGE; MEASURING INSTRUMENTS; NUCLEI; ODD-EVEN NUCLEI; RADIATION DETECTORS; RADIATIONS; RADIOISOTOPES; RESOLUTION; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; TEMPERATURE RANGE; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 5 figs., 8 refs.; http://dx.doi.org/10.11889/j.0253-3219.2016.hjs.39.100401