Shield for MRI room
- 1. Yamanashi Medical School (Japan)
Description
During the time of shielding the MRI room, it is necessary to consider shielding of the magnetic field as well as that of electromagnetic wave (radio frequency). In order to reduce these influences as much as posible, we must ensure that there is sufficient distance or install an appropriate shield. This is a report of the results of our experiments which were carried out in the MRI room to evaluate the shielding effect of the magnetic field and electromagnetic waves. Materials used for the shield were a 0.6 mm MT steel plate and 0.036 mm electric copper lief. The method of measurement was proportionate to MIL-STD-285. The effect of the magnetic shield was a 1/50 reduction in the magnetic field intensity, with almost complete shielding of the electromagnetic waves. (author)
Additional details
Publishing Information
- Journal Title
- Nippon Hoshasen Gijutsu Gakkai Zasshi
- Journal Volume
- 42
- Journal Issue
- 7
- Series
- Nippon Hoshasen Gijutsu Gakkai Zasshi.
- Journal Page Range
- 973-978
- ISSN
- 0369-4305
- CODEN
- NIPHA
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 18076344
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ELECTROMAGNETIC FIELDS; ELECTROMAGNETIC RADIATION; LABORATORY SYSTEM; NMR IMAGING; NUCLEAR MAGNETIC RESONANCE; SHIELDING
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; MAGNETIC RESONANCE; RADIATIONS; RESONANCE