Published December 1988
| Version v1
Journal article
Comparison of measured responses in two spectrally-sensitive X-ray detectors to predictions obtained using the its radiation transport code
- 1. Technical Support Corp., Albuquerque, NM (USA)
- 2. Sandia National Labs., Albuquerque, NM (USA)
Description
Responses of a Ge detector and a filtered TLD array detector have been measured at a steady-state bremsstrahlung source (the Pelletron), at endpoint energies from 150 to 900 keV. Predictions of detector response using Monte Carlo ITS codes are found to be in excellent agreement with measured responses for both detectors. These results extend the range of validity of the ITS codes. With calibration provided by these experiments and by ITS predictions, dose-depth data from the TLD arrays can be used to estimate flash X-ray source endpoint energies
Additional details
Publishing Information
- Journal Title
- IEEE Transactions on Nuclear Science
- Journal Volume
- 35
- Journal Issue
- 6
- Series
- IEEE Trans. Nucl. Sci.
- Journal Page Range
- 1272-1277
- ISSN
- 0018-9499
- CODEN
- IETNA
Conference
- Title
- 25. annual conference on nuclear and space radiation effects.
- Dates
- 12-15 Jul 1988.
- Place
- Portland, OR (USA).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 20057567
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- BREMSSTRAHLUNG; CALIBRATION; COMPUTER CODES; DEPTH DOSE DISTRIBUTIONS; GE SEMICONDUCTOR DETECTORS; I CODES; MONTE CARLO METHOD; RADIATION DETECTORS; RADIATION TRANSPORT; X RADIATION
- Descriptors DEC
- DISTRIBUTION; ELECTROMAGNETIC RADIATION; IONIZING RADIATIONS; MEASURING INSTRUMENTS; RADIATION DOSE DISTRIBUTIONS; RADIATIONS; SEMICONDUCTOR DETECTORS; SPATIAL DISTRIBUTION; SPATIAL DOSE DISTRIBUTIONS
Optional Information
- Secondary number(s)
- CONF-880730--.