Three-dimensional beam scanning for proton therapy
Creators
- 1. National Inst. of Radiological Sciences, Chiba (Japan). Div. of Physics
- 2. Tsukuba Univ., Sakura, Ibaraki (Japan). Inst. of Basic Medical Sciences
Description
An important goal of radiation therapy is to localize the dose in the target volume. The favorable properties of proton beams, such as a well-determined range and straight penetration through tissues, have been used to develop a three-dimensional beam scanning method. This system consists of a two-dimensional beam scanning system and an energy degrader of variable thickness which are both controlled by a beam monitor that is connected to a minicomputer. A typically conformed dose distribution with the three-dimensional beam scanning system was measured by a specially designed multiwire ionization chamber. The results have shown that the system works satisfactorily and considerably reduces the radiation dose outside the target volume. (orig.)
Additional details
Publishing Information
- Journal Title
- Nucl. Instrum. Methods Phys. Res.
- Journal Volume
- 214
- Journal Issue
- 2/3
- Series
- Nucl. Instrum. Methods Phys. Res.
- Journal Page Range
- 491-496
- ISSN
- 0167-5087
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 15009945
- Subject category
- S43: PARTICLE ACCELERATORS;
- Descriptors DEI
- BEAM MONITORS; BEAM TRANSPORT; COLLIMATORS; COMPUTER CODES; MICROPROCESSORS; PROTON BEAMS; RADIOTHERAPY; REMOTE CONTROL; SPATIAL DOSE DISTRIBUTIONS; THREE-DIMENSIONAL CALCULATIONS
- Descriptors DEC
- BEAMS; CONTROL; DISTRIBUTION; ELECTRONIC CIRCUITS; MEASURING INSTRUMENTS; MEDICINE; MICROELECTRONIC CIRCUITS; NUCLEON BEAMS; PARTICLE BEAMS; RADIATION DOSE DISTRIBUTIONS; SPATIAL DISTRIBUTION; THERAPY