Performance evaluation of a small animal PET scanner. Spatial resolution characterization using 18F and 11C

  • 1. Medical Physics Department, Policlinico S. Orsola-Malpighi, Via Massarenti 9, Bologna 40138 (Italy)
  • 2. Nuclear Medicine Department, Policlinico S. Orsola-Malpighi, Via Massarenti 9, Bologna 40138 (Italy)
  • 3. Dipartimento di Fisica, Universita degli studi di Bologna, Viale Berti-Pichat 6, Bologna 40136 (Italy)

Description

The performance of a depth-of-interaction preclinical small animal PET scanner (GE eXplore Vista) has been investigated. To this purpose several parameters such as spatial resolution, sensitivity, scatter fraction and noise equivalent counts were measured. Because of the different mean positron range of 18F and 11C attention was focused on measuring the effect of such difference on the scanner spatial resolution. The spatial resolution was measured using a point, a line source and a high resolution phantom. Scatter fraction and noise equivalent counts were measured using mouse and rat size objects. Results show that the eXplore Vista provides a spatial resolution of 1.48 mm and sensitivity up to 6%. In vivo images of small mice were also presented

Additional details

Identifiers

DOI
10.1016/j.nima.2006.10.066;
PII
S0168-9002(06)01863-8;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
571
Journal Issue
1-2
Journal Page Range
p. 215-218
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
1. international conference on molecular imaging technology
Acronym
EuroMedIm 2006
Dates
9-12 May 2006
Place
Marseille (France)

Optional Information

Copyright
Copyright (c) 2006 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.