Published May 1989 | Version v1
Journal article

New commercial method for the enzymatic determination of creatinine in serum and urine evaluated: Comparison with a kinetic Jaffe method and isotope dilution-mass spectrometry

  • 1. Huddinge Univ. Hospital (Sweden)

Description

We evaluated a new, simple, enzymatic kinetic method from Wako Chemicals GmbH in comparison with a kinetic Jaffe method by using isotope dilution-mass spectrometry (ID-MS) as a reference method. An ID-MS-calibrated serum standard was used. Both the enzymatic and the Jaffe method correlated well with ID-MS, except for sera with high concentrations of bilirubin. Ethyl acetoacetate, acetone, and glucose in serum interfered somewhat with the Jaffe method but not with the enzymatic method. We conclude that the present enzymatic method has merit as compared with a Jaffe method for routine work, but is more expensive

Additional details

Publishing Information

Journal Title
Clinical Chemistry (Winston-Salem, N.C.)
Journal Volume
35
Journal Issue
5
Series
Clin. Chem. (Winston-Salem, N.C.).
Journal Page Range
835-837
ISSN
0009-9147
CODEN
CLCHA