Published February 1992 | Version v1
Report Open

Calibration and evaluation of neutron moisturemeter

  • 1. Inst. For Application of Atomic Energy, CAAS (China)

Description

Factors influencing the calibration curve of neutron moisture meter, such as soil type, texture, volume weight and depth, were studied. When the soil bulk density water content is between 15% to 45%, the calibration curve is approximately a straight line, and the intercept and slope are only influenced by the above factors. The growing plants also influence the calibration curve slightly. The measuring error for top soil (< 20 cm) is larger. The relative error between neutron method and weighing method is about 8%. The neutron method has many advantages such as non-interfering, simple, fast and non-time-delay

Availability note (English)

MF available from INIS under the Report Number.

Files

24011973.pdf

Files (448.0 kB)

Name Size Download all
md5:9554b21eacdfca3f0f1db487822e1339
448.0 kB Preview Download

Additional details

Publishing Information

ISBN
7-5022-0661-5
Imprint Pagination
14 p.
Report number
CNIC--00591

Optional Information

Secondary number(s)
CSNAS--0054.