Published February 1992
| Version v1
Report
Open
Calibration and evaluation of neutron moisturemeter
Creators
- 1. Inst. For Application of Atomic Energy, CAAS (China)
Description
Factors influencing the calibration curve of neutron moisture meter, such as soil type, texture, volume weight and depth, were studied. When the soil bulk density water content is between 15% to 45%, the calibration curve is approximately a straight line, and the intercept and slope are only influenced by the above factors. The growing plants also influence the calibration curve slightly. The measuring error for top soil (< 20 cm) is larger. The relative error between neutron method and weighing method is about 8%. The neutron method has many advantages such as non-interfering, simple, fast and non-time-delay
Availability note (English)
MF available from INIS under the Report Number.Files
24011973.pdf
Files
(448.0 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:9554b21eacdfca3f0f1db487822e1339
|
448.0 kB | Preview Download |
Additional details
Publishing Information
- ISBN
- 7-5022-0661-5
- Imprint Pagination
- 14 p.
- Report number
- CNIC--00591
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 24011973
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S58: GEOSCIENCES;
- Descriptors DEI
- CALIBRATION; COMPARATIVE EVALUATIONS; DATA COVARIANCES; MOISTURE GAGES; NEUTRONS; RADIOMETRIC GAGES; SOILS
- Descriptors DEC
- BARYONS; ELEMENTARY PARTICLES; EVALUATION; FERMIONS; HADRONS; MEASURING INSTRUMENTS; NUCLEONS
Optional Information
- Secondary number(s)
- CSNAS--0054.