Signal form in tower-type ionization calorimeters
- 1. Gosudarstvennyj Komitet po Ispol'zovaniyu Atomnoj Ehnergii SSSR, Serpukhov (Russian Federation). Inst. Fiziki Vysokikh Ehnergij
- 2. Joint Inst. for Nuclear Research, Dubna (Russian Federation)
Description
The electric model of ionization calorimeter tower with 20x20 cm section and 6 nuclear length longitudinal size along the tower structure is shown to be 160 ns, the tower characteristic impedance is 5 Ohm, the signal beeing significantly influenced by dispersion. The high counting rate operation of the calorimeter considered, requires perfect matching of the detector characteristic impedance with the amplifier input impedance. 7 refs.; 9 figs
Availability note (English)
MF available from INIS under the Report Number.Files
24059740.pdf
Files
(420.2 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:1a38c13d614994c3a94fa2a07dc3decf
|
420.2 kB | Preview Download |
System files
(36.1 kB)
| Name | Size | Download all |
|---|
Additional details
Additional titles
- Original title (Russian)
- Форма сигнала в ионизационных калориметрах башенного типа
Publishing Information
- Imprint Pagination
- 14 p.
- Report number
- IFVE-OEIUNK--91-16
INIS
- Country of Publication
- Russian Federation
- Country of Input or Organization
- Russian Federation
- INIS RN
- 24059740
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COUNTING RATES; ELECTRIC IMPEDANCE; EQUIVALENT CIRCUITS; IONIZATION CHAMBERS; PULSE CIRCUITS; SHOWER COUNTERS
- Descriptors DEC
- ELECTRONIC CIRCUITS; IMPEDANCE; MEASURING INSTRUMENTS; RADIATION DETECTORS
Optional Information
- Notes
- Submitted to NIM.