Signal form in tower-type ionization calorimeters

  • 1. Gosudarstvennyj Komitet po Ispol'zovaniyu Atomnoj Ehnergii SSSR, Serpukhov (Russian Federation). Inst. Fiziki Vysokikh Ehnergij
  • 2. Joint Inst. for Nuclear Research, Dubna (Russian Federation)

Description

The electric model of ionization calorimeter tower with 20x20 cm section and 6 nuclear length longitudinal size along the tower structure is shown to be 160 ns, the tower characteristic impedance is 5 Ohm, the signal beeing significantly influenced by dispersion. The high counting rate operation of the calorimeter considered, requires perfect matching of the detector characteristic impedance with the amplifier input impedance. 7 refs.; 9 figs

Availability note (English)

MF available from INIS under the Report Number.

Files

24059740.pdf

Files (420.2 kB)

Name Size Download all
md5:1a38c13d614994c3a94fa2a07dc3decf
420.2 kB Preview Download

System files (36.1 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Форма сигнала в ионизационных калориметрах башенного типа

Publishing Information

Imprint Pagination
14 p.
Report number
IFVE-OEIUNK--91-16

INIS

Country of Publication
Russian Federation
Country of Input or Organization
Russian Federation
INIS RN
24059740
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
COUNTING RATES; ELECTRIC IMPEDANCE; EQUIVALENT CIRCUITS; IONIZATION CHAMBERS; PULSE CIRCUITS; SHOWER COUNTERS
Descriptors DEC
ELECTRONIC CIRCUITS; IMPEDANCE; MEASURING INSTRUMENTS; RADIATION DETECTORS

Optional Information

Notes
Submitted to NIM.