Published November 2001
| Version v1
Journal article
A study on head dose in computed tomography
- 1. Suzhou Univ. (China). Nuclear Medicine College
- 2. City University of Hong Kong (China)
- 3. Suzhou First People's Hospital, Suzhou (China)
Description
The dose profiles at the center and the places 1 cm away from the surface of a standard dosimetry phantom of a head were measured with thermoluminescent dosimeters (TLD) by using a Hitachi W-1000 CT system. Regression equations for the scattered radiation dose as a function of the distance from the radiation center were obtained by fitting the measured data. The CT dose index (CT-DI) was different from those computed by integrating over different intervals. Some problems worthy of special attention were discussed. This study provides a technical base for CT dose monitoring and promoting application of International Basic Standard of Radiation Protection and The Safety of Radiation Source (IBSS) guide level
Additional details
Publishing Information
- Journal Title
- Nuclear Techniques
- Journal Volume
- 24
- Journal Issue
- 11
- Journal Page Range
- p. 941-945
- ISSN
- 0253-3219
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 33010955
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY;
- Resource subtype / Literary indicator
- Numerical Data
- Descriptors DEI
- COMPUTERIZED TOMOGRAPHY; ELECTRON SCANNING; EXPERIMENTAL DATA; HEAD; MEDICAL EXAMINATIONS; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIATION MONITORING; REGRESSION ANALYSIS; SAFETY STANDARDS; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- BODY; DATA; DIAGNOSTIC TECHNIQUES; DOSEMETERS; INFORMATION; LUMINESCENT DOSEMETERS; MATHEMATICS; MEASURING INSTRUMENTS; MEDICAL SURVEILLANCE; MOCKUP; MONITORING; NUMERICAL DATA; STANDARDS; STATISTICS; STRUCTURAL MODELS; TOMOGRAPHY