Published December 1995 | Version v1
Journal article

Changes to dose in the build-up region when using multi-leaf collimators in place of lead blocks supported on an accessory tray

Description

Doses in the build-up region have been compared for regular fields with and without a perspex accessory tray and for two irregular fields defined by either a Philips multi-leaf collimator (MLC) or lead blocks. The results show an increase in doses within the build-up region, by up to a factor of two, primarily due to the presence of the accessory tray. Fields shaped with the MLC exhibited build-up characteristics similar to those in unblocked fields and hence there will be systematic changes to the build-up dose when treatments are transferred from lead blocking techniques to those using an MLC mounted within the treatment head

Additional details

Identifiers

PII
0167814095016627;

Publishing Information

Journal Title
Radiotherapy and Oncology
Journal Volume
37
Journal Issue
3
Journal Page Range
p. 225-229
ISSN
0167-8140
CODEN
RAONDT

INIS

Country of Publication
Ireland
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
34044457
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
COLLIMATORS; LEAD; RADIATION DOSES; SHIELDING; THERAPEUTIC USES
Descriptors DEC
DOSES; ELEMENTS; METALS; USES

Optional Information

Copyright
Copyright (c) 1995 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.