Published December 1995
| Version v1
Journal article
Changes to dose in the build-up region when using multi-leaf collimators in place of lead blocks supported on an accessory tray
Creators
Description
Doses in the build-up region have been compared for regular fields with and without a perspex accessory tray and for two irregular fields defined by either a Philips multi-leaf collimator (MLC) or lead blocks. The results show an increase in doses within the build-up region, by up to a factor of two, primarily due to the presence of the accessory tray. Fields shaped with the MLC exhibited build-up characteristics similar to those in unblocked fields and hence there will be systematic changes to the build-up dose when treatments are transferred from lead blocking techniques to those using an MLC mounted within the treatment head
Additional details
Identifiers
- PII
- 0167814095016627;
Publishing Information
- Journal Title
- Radiotherapy and Oncology
- Journal Volume
- 37
- Journal Issue
- 3
- Journal Page Range
- p. 225-229
- ISSN
- 0167-8140
- CODEN
- RAONDT
INIS
- Country of Publication
- Ireland
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 34044457
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- COLLIMATORS; LEAD; RADIATION DOSES; SHIELDING; THERAPEUTIC USES
- Descriptors DEC
- DOSES; ELEMENTS; METALS; USES
Optional Information
- Copyright
- Copyright (c) 1995 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.