Lead iodide optical detectors for gamma ray spectroscopy

  • 1. RMD, Inc., Watertown, MA (United States)

Description

This paper describes the research performed in developing low noise, high quantum efficiency lead iodide photodetectors for use with scintillators. These photodetectors operate with very low leakage current and show high quantum efficiency (>60%) in 350 to 500 mn region. Successful scintillation studies have been performed at room temperature as well as elevated temperatures (100 degrees C) using PbI2 photodetectors with LSO and CsI(Na) scintillators. Detailed analysis of noise has also been performed and potential applications are discussed

Additional details

Publishing Information

Publisher
IEEE Service Center.
Imprint Place
Piscataway, NJ (United States)
Imprint Title
1996 IEEE nuclear science symposium - conference record. Volumes 1, 2 and 3
Imprint Pagination
2138 p.
Journal Page Range
p. 21-24.

Conference

Title
Institute of Electrical and Electronic Engineers (IEEE) nuclear science symposium and medical imaging conference.
Dates
2-9 Nov 1996.
Place
Anaheim, CA (United States).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
28068817
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
GAMMA SPECTROMETERS; LEAD IODIDES; PHOTODETECTORS; QUANTUM EFFICIENCY
Descriptors DEC
EFFICIENCY; HALIDES; HALOGEN COMPOUNDS; IODIDES; IODINE COMPOUNDS; LEAD COMPOUNDS; LEAD HALIDES; MEASURING INSTRUMENTS; SPECTROMETERS

Optional Information

Secondary number(s)
CONF-961123--.