Assessment of myocardial perfusion with MRI using a modified dual bolus method
Creators
- 1. Department of Clinical Radiology, Kuopio University Hospital, Kuopio (Finland)
- 2. Heart Center, Kuopio University Hospital, Kuopio (Finland)
- 3. Institute of Clinical Medicine, University of Eastern Finland, Kuopio (Finland)
- 4. Department of Applied Physics, University of Eastern Finland, Kuopio (Finland)
- 5. Department of Clinical Physiology and Nuclear Medicine, Kuopio University Hospital, Kuopio (Finland)
Description
Quantification of regional myocardial blood flow (rMBF) with first-pass magnetic resonance imaging (FP-MRI) requires two contrast agent injections (dual bolus technique), inducing error in the determined rMBF if the injections differ. We hypothesize that using input and residue curves of the same injection would be more reliable. We aim to introduce and evaluate a novel method to correct the high concentration arterial input function (AIF) for determination of rMBF. Sixteen patients with non-Hodgkin's lymphoma were examined before and after chemotherapy. The input function was solved by correcting initial high concentration AIF using the ratio of low and high contrast AIF areas, normalized by corresponding heart rates (modified dual bolus method). For comparison, the scaled low contrast AIF was used as an input function (dual bolus method). Unidirectional transfer coefficient Ktrans was calculated using both methods. Ktrans-values determined with the dual bolus (0.81 ± 0.32 ml g−1 min−1) and modified dual bolus (0.77 ± 0.42 ml g−1 min−1) methods were in agreement (p = 0.21). Mean Ktrans-values increased from 0.76 ± 0.43 to 0.89 ± 0.55 ml g−1 min−1 after chemotherapy (p = 0.17). The modified dual bolus technique agrees with the dual bolus technique (R2 = 0.899) when the low and high contrast injections are similar. However, when this is not the case, the modified dual bolus technique may be more reliable. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/0967-3334/35/4/533Additional details
Identifiers
Publishing Information
- Journal Title
- Physiological Measurement (Print)
- Journal Volume
- 35
- Journal Issue
- 4
- Journal Page Range
- p. 533-547
- ISSN
- 0967-3334
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 47054958
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S75: CONDENSED MATTER PHYSICS, SUPERCONDUCTIVITY AND SUPERFLUIDITY;
- Descriptors DEI
- BLOOD FLOW; CHEMOTHERAPY; COMPARATIVE EVALUATIONS; CONCENTRATION RATIO; DIAGRAMS; ERRORS; HEART; INJECTION; LYMPHOMAS; NMR IMAGING; PATIENTS
- Descriptors DEC
- BODY; CARDIOVASCULAR SYSTEM; DIAGNOSTIC TECHNIQUES; DIMENSIONLESS NUMBERS; DISEASES; EVALUATION; IMMUNE SYSTEM DISEASES; INFORMATION; INTAKE; MEDICINE; NEOPLASMS; ORGANS; THERAPY