Published October 1994
| Version v1
Report
Results on substituting the NaI by a Ge detector in a simple shadow shield whole body counter
Creators
- 1. Nuclear Engineering and Analytics Inc., Dresden (Germany)
- 2. Amersham Buchler Ltd. ampersand Co., Gieselweg (Germany)
Description
Since 1976 a whole body counter (WBC) has been used at the Rossendorf Research Centre for measuring the internal contamination of workers. The HP Ge detector (n-type, efficiency 43%) was substituted for a 3 x 3 NaI(Tl) detector. From the simple substitution, without any modification of the WBC construction the following improvements have been achieved: (1) Clear nuclide identification; (2) Decrease of the lower limit of detection from about 100 Bq to about 60 Bq, or reduction of the measuring time from 2000 s to 1000 s; (3) Faster availability of the analysis results by using a commercially available spectrum analysis software package; and (4) Simpler calibration of the WBC due to the use of a multiline emitter
Additional details
Publishing Information
- Imprint Title
- Fourth conference on radiation protection and dosimetry: Proceedings, program, and abstracts
- Imprint Pagination
- 620 p.
- Journal Page Range
- p. 327-335.
- Report number
- ORNL/TM--12817
Conference
- Title
- 43. Franz-Pacher colloquium on geomechanics.
- Dates
- 13-14 Oct 1994.
- Place
- Salzburg (Austria).
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 26046958
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- CALIBRATION; COMPARATIVE EVALUATIONS; EFFICIENCY; FEASIBILITY STUDIES; HIGH-PURITY GE DETECTORS; PERFORMANCE TESTING; PERSONNEL DOSIMETRY; PERSONNEL MONITORING; RADIATION PROTECTION; WHOLE-BODY COUNTING
- Descriptors DEC
- COUNTING TECHNIQUES; DOSIMETRY; EVALUATION; GE SEMICONDUCTOR DETECTORS; MEASURING INSTRUMENTS; MONITORING; RADIATION DETECTORS; RADIATION MONITORING; SEMICONDUCTOR DETECTORS; TESTING
Optional Information
- Secondary number(s)
- CONF-9410171--.