Published 1987 | Version v1
Report Open

Possibility of improvement of spatial resolution in multistep avalanche chambers

  • 1. AN Tadzhikskoj SSR, Dushanbe. Fiziko-Tekhnicheskij Inst.
  • 2. Joint Inst. for Nuclear Research, Dubna (USSR). Lab. of High Energy

Description

The influence of the signal amplitude on the spatial resolution of multistep avalanche chambers (MSAC) has been determined for the detection of β-particles. Large signal amplitudes (25% of events) deteriorate substantially the spatial resolution of the detectors. The analysis of the signal amplitude with selective coordinate detection allows the spatial resolution of MSAC to be improved approximately by a factor of 2. It can be equal to 0.5-1 mm in the detection of the isotropic radiation of 14C and 32P

Files

19041947.pdf

Files (122.7 kB)

Name Size Download all
md5:3a8394eaeed4fc1e097bcd2a06afd744
122.7 kB Preview Download

System files (12.9 kB)

Name Size Download all

Additional details

Additional titles

Original title (Russian)
Возможност' улучшения пространственного разрешения в многоступенчатых лавинных камерах

Publishing Information

Imprint Pagination
4 p.
Report number
JINR--13-87-290

Optional Information

Notes
7 refs.; 3 figs.