Measurements of optical constants at terahertz--or T-ray--frequencies have been performed extensively using terahertz time-domain spectroscopy (THz-TDS). Spectrometers, together with physical models explaining the interaction between a sample and T-ray radiation, are progressively being developed. Nevertheless, measurement errors in the optical constants, so far, have not been systematically analyzed. This situation calls for a comprehensive analysis of measurement uncertainty in THz-TDS systems. The sources of error existing in a terahertz spectrometer and throughout the parameter estimation process are identified. The analysis herein quantifies the impact of each source on the output optical constants. The resulting analytical model is evaluated against experimental THz-TDS data

Additional details

Identifiers

Publishing Information

Journal Title
Journal of the Optical Society of America. Part B, Optical Physics
Journal Volume
25
Journal Issue
6
Journal Page Range
p. 1059-1072
ISSN
0740-3224
CODEN
JOBPDE

INIS

Country of Publication
United States
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
39110394
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
ERRORS; MEASURING METHODS; SPECTROMETERS; SPECTROSCOPY; THZ RANGE
Descriptors DEC
FREQUENCY RANGE; MEASURING INSTRUMENTS