Published December 1976 | Version v1
Journal article

Specification and calibration of a quadrupole mass filter

  • 1. Ibaraki Univ., Hitachi (Japan). Faculty of Engineering

Description

Specification and calibration of a quadrupole mass filter for the study of mass separation phenomena are described. The mass filter is operated in a range of mass number from 1 to 100a.m.u. with a resolution of 50. Analyzing rods are 150mm in length and are energized by HF potential of 1 MHz. Sensitivity is of the order of 1.0x10-7 A/Torr for N2 gas. (auth.)

Additional details

Publishing Information

Journal Title
Ibaraki Daigaku Kogakubu Kenkyu Shuho
Journal Volume
24
Series
Ibaraki Daigaku Kogakubu Kenkyu Shuho.
Journal Page Range
185-188

INIS

Country of Publication
Japan
Country of Input or Organization
Japan
INIS RN
8341142
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
CALIBRATION; CONTROL SYSTEMS; EFFICIENCY; FILTERS; MASS; MASS SPECTROMETERS; QUADRUPOLES; RESOLUTION; SENSITIVITY; SPECIFICATIONS
Descriptors DEC
MEASURING INSTRUMENTS; MULTIPOLES; SPECTROMETERS