Published March 1978
| Version v1
Journal article
Measurement of scintillation decay curves by a single photon counting technique
Description
An improved apparatus suitable for the measurement of spectroscopic scintillation decay curves has been developed by combination of a single photon counting technique and a delayed coincidence method. The time resolution of the apparatus is improved up to 1.16 nsec (FWHM), which is obtained from the resolution function of the system for very weak Cherenkov light flashes. Systematic measurement of scintillation decay curves is made for liquid and crystal scintillators including PPO-toluene, PBD-xylene, PPO-POPOP-toluene, anthracene and stilbene. (auth.)
Additional details
Publishing Information
- Journal Title
- Denshi Gijutsu Sogo Kenkyusho Iho
- Journal Volume
- 42
- Journal Issue
- 1-2
- Series
- Denshi Gijutsu Sogo Kenkyusho Iho.
- Journal Page Range
- 55-67
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 10446077
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ACCURACY; ANTHRACENE; COINCIDENCE METHODS; PULSES; RADIOLUMINESCENCE; SCINTILLATION COUNTING; SCINTILLATIONS; STILBENE; TIME RESOLUTION; TOLUENE; XYLENES
- Descriptors DEC
- AROMATICS; CONDENSED AROMATICS; COUNTING TECHNIQUES; HYDROCARBONS; LUMINESCENCE; ORGANIC COMPOUNDS; RESOLUTION; TIMING PROPERTIES