Published 1996 | Version v1
Journal article

Comparison of LiF:Mg,Cu,P (MCP-N, GR-200A) and α-Al2O3:C TL detectors in short-term measurements of natural radiation

  • 1. Institute of Nuclear Physics, Krakow (Poland)
  • 2. Risoe National Lab., Roskilde (Denmark)
  • 3. Centro de Investigaciones Energeticas, Medioambientales y Tecnologicas (CIEMAT), Madrid (Spain)
  • 4. Centre of Oncology, Krakow (Poland)

Description

Results are reported of tests of GR-200A (LiF:Mg,Cu,P Chinese production), MCP-N (LiF:Mg,Cu,P Polish production) and α-Al2O3 highly sensitive TL detectors used in short-term measurements of natural radiation performed within the EU Intercalibration by the Polish (INP) and Spanish (CIEMAT) participants. The experiment was carried out in Roskilde, Denmark, by Riso National Laboratory. To assess only the cosmic component of natural radiation, detectors were exposed on a wooden pier. The actual dose rate was monitored over the total exposure time of three days by a Reuter-Stokes high pressure ionisation chamber. Results obtained with MCP-N and GR-200A were, for both teams, about 32 nGy.h-1, for α-Al2O3:C about 49 nGy.h-1, while the chamber-evaluated mean dose rate was about 40 nGy.h-1. It was found that all tested detectors (MCP-N, GR-200A and α-Al2O3:C) are capable of measuring natural gamma dose rates after an exposure of only a few hours. (Author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
66
Journal Issue
1-4
Journal Page Range
p. 157-160.
ISSN
0144-8420
CODEN
RPDODE

Conference

Title
11. international conference on solid state dosimetry.
Dates
10-14 Jul 1995.
Place
Budapest (Hungary).

Optional Information

Contract/Grant/Project number
Contract CEC F13P-CT920018
Secondary number(s)
CONF--950708.