Published December 1985 | Version v1
Journal article

Three-component superconductor magnetometer

  • 1. Novosibirsk Electrical Engineering Institute

Description

A three-component superconductor magnetometer is described. Particular attention in its design was given to reducing measurement errors due to the mutual effect of the channels, lack of identity of their transmission coefficients, and distortion of the magnetic field being measured by the superconducting shields. A superconducting connector is used in each channel. The immersed part of the three-component magnetometer is a single unit with an outside diameter of 20.5 mm. The sensitivity of the magnetometer to an external field is about 5.10-13 T. The precision of the mutually perpendicular faces of the cube with the pickup, loops within a few angular minutes. The mutual effect of the channels is ≤ 0.03%

Additional details

Publishing Information

Journal Title
Instrum. Exp. Tech. (Engl. Transl.)
Journal Volume
26
Journal Issue
3
Series
Instrum. Exp. Tech. (Engl. Transl.).
Journal Page Range
683-685
ISSN
0020-4412
CODEN
INETA

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
19020716
Subject category
S71: CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSICS;
Descriptors DEI
MAGNETIC FIELDS; MAGNETIC FLUX; MAGNETOMETERS; PERFORMANCE; SENSITIVITY; SPECIFICATIONS; SQUID DEVICES
Descriptors DEC
ELECTRONIC EQUIPMENT; EQUIPMENT; FLUXMETERS; MEASURING INSTRUMENTS; MICROWAVE EQUIPMENT; SUPERCONDUCTING DEVICES