Published February 1981 | Version v1
Journal article

Comparison of count rate parameters in gamma cameras

Description

Pulse pile-up is a fundamental problem that limits the ability of a gamma camera to produce high quality images at high count rates. Pulse pile-up results in loss of events and spatial distortions in the image. The question asked in this study is how well do some of the test procedures for measuring count rate performance in gamma cameras compare. In four gamma cameras we compared measurements of spectral fraction, count-rate curve, pulse-pair resolution, deadtime, maximum count rate, full width at half maximum of the line spread function, and misplaced event count rate. The results indicated that no one technique provides a complete description of the count-rate effects in gamma cameras. The misplaced-event measurements provided the most information

Additional details

Publishing Information

Journal Title
J. Nucl. Med.
Journal Volume
22
Journal Issue
2
Series
J. Nucl. Med.
Journal Page Range
161-168
ISSN
0022-3123

Conference

Title
28. annual meeting of Society of Nuclear Medicine.
Dates
16 Jun 1981.
Place
Las Vegas, NV, USA.

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
12607838
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE; S07: ISOTOPES AND RADIATION SOURCES;
Resource subtype / Literary indicator
Conference
Descriptors DEI
COMPARATIVE EVALUATIONS; COUNTING RATES; DEAD TIME; GAMMA CAMERAS; IMAGES; PERFORMANCE TESTING; PULSE PILEUP; RESOLUTION; SCINTISCANNING; SENSITIVITY
Descriptors DEC
CAMERAS; COUNTING TECHNIQUES; DIAGNOSTIC TECHNIQUES; RADIOISOTOPE SCANNING; TESTING; TIMING PROPERTIES