Description
A mammographic imaging set-up using silicon microstrip detectors in edge-on geometry was simulated using the GEANT package. Deposited energy in tissue of various thicknesses was evaluated and shown to agree to within 10% with reference calculations. Optimal energies as well as spectra for mammography with silicon detectors were determined by maximizing the figure of merit of a realistic imaging set-up. The scattered to primary radiation ratio was studied for various detector geometries. It was found that fan-shaped detectors are needed to maintain the image quality for divergent photon beams. (author)
Additional details
Publishing Information
- Journal Title
- Physics in Medicine and Biology (Online)
- Journal Volume
- 45
- Journal Issue
- 7
- Journal Page Range
- p. 2029-2041
- ISSN
- 1361-6560
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 43066956
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DESIGN; ENERGY ABSORPTION; IMAGE PROCESSING; MAMMARY GLANDS; OPTIMIZATION; PHOTON BEAMS; RADIATION DETECTORS; SILICON
- Descriptors DEC
- ABSORPTION; BEAMS; BODY; DIAGNOSTIC TECHNIQUES; ELEMENTS; GLANDS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; ORGANS; PROCESSING; RADIOLOGY; SEMIMETALS; SORPTION
Optional Information
- Notes
- 23 refs; This record replaces 31040287