Published July 2000 | Version v1
Journal article

Design optimization of a breast imaging system based on silicon microstrip detectors

  • 1. DESY, Notkestr. 85, D-22607 Hamburg (Germany).
  • 2. Jozef Stefan Institute, Jamova 39, SI-1000, Ljubljana (Slovenia)
  • 3. Department of Physics, University of Ljubljana, Jadranska 19, SI-1000, Ljubljana (Slovenia)

Description

A mammographic imaging set-up using silicon microstrip detectors in edge-on geometry was simulated using the GEANT package. Deposited energy in tissue of various thicknesses was evaluated and shown to agree to within 10% with reference calculations. Optimal energies as well as spectra for mammography with silicon detectors were determined by maximizing the figure of merit of a realistic imaging set-up. The scattered to primary radiation ratio was studied for various detector geometries. It was found that fan-shaped detectors are needed to maintain the image quality for divergent photon beams. (author)

Additional details

Publishing Information

Journal Title
Physics in Medicine and Biology (Online)
Journal Volume
45
Journal Issue
7
Journal Page Range
p. 2029-2041
ISSN
1361-6560

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
43066956
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; DESIGN; ENERGY ABSORPTION; IMAGE PROCESSING; MAMMARY GLANDS; OPTIMIZATION; PHOTON BEAMS; RADIATION DETECTORS; SILICON
Descriptors DEC
ABSORPTION; BEAMS; BODY; DIAGNOSTIC TECHNIQUES; ELEMENTS; GLANDS; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; ORGANS; PROCESSING; RADIOLOGY; SEMIMETALS; SORPTION

Optional Information

Notes
23 refs; This record replaces 31040287