Published August 1978
| Version v1
Journal article
Simultaneous monitoring system of thermal neutron flux for boron-neutron capture therapy
- 1. Teikyo Univ. Hospital, Tokyo
Description
A simultaneous monitoring device of neutron flux for boron-neutron capture therapy has been developed. A silicon surface-barrier detector mounted with borax is used to count neutrons. The silicon detector is 2.5 mm in diameter and is attached to a cable 1.6 mm in diameter and about 1 m in length. The relative error of the linearity between the counting rate and the neutron flux is within 5%, including statistical fluctuations. The relative error of counting rate is within 5% during 240 min of irradiation with a constant neutron flux of 1.7 x 109 n/cm2 sec
Additional details
Publishing Information
- Journal Title
- Radiat. Res.
- Journal Volume
- 75
- Journal Issue
- 2
- Series
- Radiat. Res.
- Journal Page Range
- 243-251
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 10470019
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S07: ISOTOPES AND RADIATION SOURCES;
- Descriptors DEI
- BORON; BRAIN; DOSIMETRY; ERRORS; NEOPLASMS; NEUTRON CAPTURE THERAPY; NEUTRON FLUX; RADIATION MONITORING; THERMAL NEUTRONS