Published January 1989 | Version v1
Report Open

Precision charge amplification and digitization system for a scintillating and lead glass array

Description

A 544-channel low-noise, high-rate, precision charge amplification and ADC system was constructed for the Fermilab Experiment 705 electromagnetic calorimeter, which employs SCG1-C scintillating glass and SF5 lead glass instrumented with photo-multiplier tubes. A general discussion of the system is given, and the charge amplification, fast trigger pulse generation, and analog to digital conversion aspects of the system are presented in more detail. Performance is evaluated using data from Experiment 705 and from off-line tests. Short and long term pedestal stability, baseline recovery and rate capability, linearity of response, and crosstalk between channels are discussed. 8 refs., 2 tabs

Availability note (English)

MF available from INIS under the Report Number; Available from NTIS, PC A02 - OSTI; 3 as DE89011007.

Files

20061418.pdf

Files (555.5 kB)

Name Size Download all
md5:fc9956e6bae842ceb1621036b2be1f9a
555.5 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
7 p.
Report number
FNAL/C--89/25-E

Conference

Title
IEEE nuclear science symposium.
Dates
8-10 Feb 1989.
Place
Orlando, FL (USA).

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
20061418
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Resource subtype / Literary indicator
Conference
Descriptors DEI
AMPLIFIERS; ANALOG-TO-DIGITAL CONVERTERS; ENERGY RESOLUTION; GLASS SCINTILLATORS; PERFORMANCE; SHOWER COUNTERS; SPECIFICATIONS; TRIGGER CIRCUITS
Descriptors DEC
ELECTRONIC CIRCUITS; ELECTRONIC EQUIPMENT; EQUIPMENT; MEASURING INSTRUMENTS; PHOSPHORS; PULSE CIRCUITS; RADIATION DETECTORS; RESOLUTION

Optional Information

Notes
Paper copy only, copy does not permit microfiche production.
Secondary number(s)
CONF-890297--1.