Stability of Calibration Coefficient of Type Ionization Chambers at 60Co Teletherapy Beam in Malaysia
Creators
- 1. Malaysian Nuclear Agency, Bangi, 43000 Kajang, Selangor (Malaysia)
Description
Full-text:Absorbed dose-to-water (DW) calibration is critical to the medical community to achieve the accuracy of the dose delivered to tumors during radiotherapy treatment. Therefore, the IAEA recommended annual calibration of ionization chambers with a calibration coefficient (ND,W) that must be within the acceptable range of ± 1.5%. In this study, three types ion chambers, used in most of the radiotherapy centers (RCs) have been calibrating at SSDL Laboratory of the Malaysian Nuclear Agency for seven years. The mean (μ) of the ND,W deviation together with its standard error (SE) of these chambers was calculated to measure stability of NDW. It was found that the μ±SE value lies within the acceptable limits. Thus, this indicates that the calibrated ion chambers are safely to be used for radiotherapy treatment due to the stabilization of ND,W. (author)
Availability note (English)
Available in abstract form only, full text entered in this recordAdditional details
Publishing Information
- Imprint Pagination
- 1 p.
Conference
- Title
- Nuclear Innovation and Technical Convention 2023
- Acronym
- NITC 2023
- Dates
- 24-26 Oct 2023
- Place
- Bangi (Malaysia)
INIS
- Country of Publication
- Malaysia
- Country of Input or Organization
- Malaysia
- INIS RN
- 55105413
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference, Non-conventional Literature
- Descriptors DEI
- CALIBRATION; CALIBRATION STANDARDS; COEFFICIENT OF PERFORMANCE; IONIZATION CHAMBERS; STABILITY
- Descriptors DEC
- MEASURING INSTRUMENTS; RADIATION DETECTORS; STANDARDS
Optional Information
- Notes
- Poster presentation