Dosimetric and clinical review on the application of TOMOedge mode
Creators
- 1. Dept. of Radiation Oncology, Yonsei Cancer Hospital, Seoul (Korea, Republic of)
Description
The goal of this study was to compare and analysis the dose distribution and treatment time between Tomotherapy planning with fixed jaw(FJ) and dynamic jaw(DJ). Seven patients were selected in the study including five common clinical cases(brain, head and neck(HN), lung, prostate, spine). 1) Helical Tomotherapy plans with FJ and DJ were generated with the same planning parameters such as Modulation factor, Pitch and Field width. 2) Tomoedge plans with a larger field width were generated to compare to conventional HT delivery with fixed jaw. Dosimetric evaluation indices for target coverage are Dmin, Conformity index(CI) and for whole body including target are V10%, V25%, V50%, V75% using Dose-volume histogram(DVH). Also, Treatment time and Cumulative MU were used for clinical review on Tomoedge. In case of using the same field width of Tomotherapy planning with FJ and DJ, the averaged variations were V10%: - 11.91%, V25%: -7.6%, V50% :-4.75%, V75%: -1.04%. Tomoedge with a larger field width provides the averaged variations for target coverage: Dmin: -0.72%, CI: -1.25% and also shows the tendency of a sharp Vx% decline in low dose area. The clinical improvements in the larger field width with DJ were observed in the treatment time, ranging from -51.21% to -15.11, and the Cumulative MU decrease, ranging from -57.74% to -15.31%. Conclusion : Target coverage achieved by FJ and DJ with the same field width has little differences. But integral doses on whole body efficiently decreased. Compared to the conventional HT delivery, Tomoedge with a larger field width presents a little worse target coverage. However, it provides faster treatment delivery and improved cranial-caudal target dose conformity. Therefore, Tomoedge mode is efficient in improving the treatment time and integral dose while maintaining comparable plan quality in clinic
Additional details
Publishing Information
- Journal Title
- Journal of the Korean Society for Radiotherapeutic Technology
- Journal Volume
- 26
- Journal Issue
- 2
- Series
- 10 refs, 2 figs, 4 tabs
- Journal Page Range
- p. 177-182
- ISSN
- 1598-8449
INIS
- Country of Publication
- Korea, Republic of
- Country of Input or Organization
- Korea, Republic of
- INIS RN
- 47089694
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BODY; CT-GUIDED RADIOTHERAPY; DOSIMETRY; EFFICIENCY; JAW; MAINTENANCE; PATIENTS; PLANNING; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; USES
- Descriptors DEC
- BODY; DOSES; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; RADIOTHERAPY; SKELETON; SKULL; THERAPY