Published March 1986
| Version v1
Journal article
Filter-fluorescer x-ray spectrometer using solid state detectors for γ-ray background reduction
Creators
- 1. Osaka Univ., Suita (Japan). Faculty of Engineering
Description
Filter-fluorescer x-ray spectrometer using solid state photo-detectors instead of the photomultiplier tubes in order to reduce the γ-ray background noise is reported. A significant reduction of the γ-ray background noise is expected, because solid state photo-detectors are very small in size compared with the photomultiplier tubes. It has been confirmed that the γ-ray background is reduced in the target irradiation experiments with the Gekko MII glass laser. (author)
Additional details
Publishing Information
- Journal Title
- Technol. Rep. Osaka Univ.
- Journal Volume
- 36
- Journal Issue
- 1819-1843
- Series
- Technol. Rep. Osaka Univ.
- Journal Page Range
- 147-153
- ISSN
- 0030-6177
- CODEN
- TROUA
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 18010379
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ABSORPTION; BACKGROUND NOISE; FILTERS; FLUORESCENCE; GAMMA RADIATION; HARD X RADIATION; LASER-PRODUCED PLASMA; PHOTODIODES; PLASMA DIAGNOSTICS; X-RAY SPECTRA; X-RAY SPECTROMETERS
- Descriptors DEC
- ELECTROMAGNETIC RADIATION; EMISSION; IONIZING RADIATIONS; LUMINESCENCE; MEASURING INSTRUMENTS; NOISE; PHOTON EMISSION; PLASMA; RADIATIONS; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES; SPECTRA; SPECTROMETERS; X RADIATION